16 SeriesMass and VolumetricPrecision Gas Flow MeterOperating ManualInnovative Flow and Pressure Solutions

Pagina 2

874621DB-9 Serial Port55----------Ground--------------------------------------Ground----------83----------Transmit------------------------------------

Pagina 3

9An optional industrial connector is also available:165432Pin Function Cable Color1Power In ( + )Red2 RS-232 Output Blue3 RS-232 Input Signal White4

Pagina 4 - Advice for Device Operation

10M Series Mass Flow Meter OperationThe M Series Mass Flow Meter provides a multitude of useful ow data in one simple, rugged device. The M Series ca

Pagina 5

11Gas Absolute Pressure: The M Series ow meters utilize an absolute pressure sensor to measure the line pressure of the gas ow being monitored. This

Pagina 6

12Select Menu ModePushing “Mode” once will bring up the “Select Menu” display. Push the button nearest your selection to go to the corresponding scree

Pagina 7 - Installation

13Communication SelectThe Communication Select mode is accessed by pressing the button below “Comm. RS-232” on the Select Menu display. The screen wil

Pagina 8 - AC/DC Adapter Jack

14Manufacturer Data“Manufacturer Data” is accessed by pressing the “Mfg. Data” button on the Select Menu display (Figure 7). The “Mfg 1”

Pagina 9

15Select Main>LCD Contrast(10) . 10PVM DBand (0.5% FS)PRESS Avg (008) . . . 008FLOW Avg (100) . . . . 100UP DOWN MiscFigur

Pagina 10 - 8 Pin Mini-DIN Port

16V Series Volumetric Flow Meter OperationThe V Series can have several screen “modes” depending on how the device is ordered. All V Series meters ha

Pagina 11 - DB15 Pin-out Diagrams:

17Flashing Error Message: Our volumetric ow meters and controllers display an error message (VOV = volumetric overrange) when a measured

Pagina 12 -

01/20/2009 Rev.12 DOC-ALIMAN16Conformity / Supplemental Information:The product complies with the requirements of the Low Voltage Di

Pagina 13

18In Polling Mode, the screen should be blank except the blinking cursor. In order to get the data streaming to the screen, hit the “Enter” key s

Pagina 14 - PgUP PgDWN Main

19Gas Select – The selected gas can be changed via RS-232 input. To change the selected gas, enter the following commands: In Streaming Mode: $$#<

Pagina 15

20Collecting Data:The RS-232 output updates to the screen many times per second. Very short-term events can be captured simply by disco

Pagina 16

21Sending a Simple Script File to HyperTerminal®It is sometimes desirable to capture data for an extended period of time. Standard streaming mode info

Pagina 17

22Operating PrincipleAll M and V Series Gas Flow Meters are based on the accurate measurement of volumetric ow. The volumetric ow rate is determ

Pagina 18

23Other Gases: M Series Flow Meters can easily be used to measure the ow rate of gases other than those listed as long as “non-corrosive” gas compati

Pagina 19 - RS-232 Output and Input

24Once the corrected mass ow rate at standard conditions has been determined and the density at standard conditions is known (see the densi

Pagina 20

25This reduces to: Pa Va / Za Ta = Ps Vs / Zs Ts , eliminating R and n.Alicat mass ow meters model gas ows based upon the non-ideal gas characterist

Pagina 21

26GasNumberShort Form Long FormViscosity*25 deg C 14.696 PSIADensity**25 deg C 14.696 PSIACompressibility 25 deg C 14.696 PSIA0 Air Air 184.918 1.1840

Pagina 22

27GasNumberShort Form Long FormViscosity*0 deg C 14.696 PSIADensity**0 deg C 14.696 PSIACompressibility 0 deg C 14.696 PSIA0 Air Air 172.588 1.2927 0.

Pagina 23

Protect your Warranty and Extend the Life of your ProductFailure to follow general safety and operating procedures as presented in this document and t

Pagina 24

28Volumetric Flow Meters Under PressureV Series Volumetric Flow Meters are intended for use in low pressure applications. This is because an accurate

Pagina 25 - Volume Flow vs. Mass Flow

29TROUBLESHOOTINGDisplay does not come on or is weak.Check power and ground connections.Flow reading is approximately xed either near zero or near f

Pagina 26

30RS-232 Serial Communications is not responding.Check that your meter is powered and connected properly. Be sure that the port on the computer to whi

Pagina 27

31Dimensional Drawings: page 35SpecicationMassMeterVolumetricMeterDescriptionAccuracy ± (0.8% of Reading + 0.2% of Full Scale)At calibration conditio

Pagina 28 - (NIST REFPROP 7 database)

32Dimensional Drawings: page 35, 36SpecicationMassMeterVolumetricMeterDescriptionAccuracy ± (0.8% of Reading + 0.2% of Full Scale)At calibration cond

Pagina 29

33Dimensional Drawings: page 36, 37SpecicationMassMeterVolumetricMeterDescriptionAccuracy ± (0.8% of Reading + 0.2% of Full Scale)At calibration cond

Pagina 30

34Dimensional Drawings: page 37SpecicationMassMeterVolumetricMeterDescriptionAccuracy ± (0.8% of Reading + 0.2% of Full Scale)At calibration conditio

Pagina 31 - TROUBLESHOOTING

35M & V Series:0 – 0.5SCCM0 – 1SCCM0 – 2SCCM0 – 5SCCM0 – 10SCCM0 – 20SCCM0 – 50SCCMM & V Series:0 – 100SCCM0 – 200SCCM0 – 500SCCM0 – 1SLPM0 –

Pagina 32 - Maintenance and Recalibration

36M & V Series:0 – 20SLPMM & V Series:0 – 50SLPM0 – 100SLPM

Pagina 33 - Dimensional Drawings: page 35

37M & V Series:0 – 250SLPMM & V Series:0 – 500SLPM0 – 1000SLPM0 – 1500SLPM

Pagina 34 - Mechanical Specications

Advice for Device OperationDo take note of what gas calibration setting is selected on mass and volumetric devices.Do consider that changing or mixing

Pagina 35

38Option: Totalizing Mode ScreenM and V Series Flow Meters and Controllers can be purchased with the Totalizing Mode option. This option adds an addit

Pagina 36 - Dimensional Drawings: page 37

39Alicat Portable Meters and GaugesAlicat Portable Flow Meters and Gauges use a common 9 Volt battery located in the top section of your meter. Output

Pagina 37 - Alicat Scientific, Inc

40Option: Remote Electronics for High Line or Gas TemperaturesSome applications involve operating temperatures outside the standard Alicat device spec

Pagina 38

41Accessory: Flow Vision™ SC SoftwareFlow Vision™ SC is an intuitive software interface to help your test cycles run smoother and shorten your enginee

Pagina 39

42AccessoriesPart Number DescriptionFLOWVISIONSC Flow Vision™ SC software for interface with all Alicat instrumentsBB-9 Multi-Drop BoxPCASE Industrial

Pagina 40

43DB15 Pin-OutNote: The above pin-out is correct for units with a DB15 pin-out. If your unit is marked DB15H or DB15K, you must use the correct pin-ou

Pagina 41 - Replacing the Battery:

44DB15H Pin-OutPin Number Function6 Signal Out7 Ground11 Supply14 Set-Point10 Secondary Out2 RS-232 Receive15 RS-232 Send711 14151026DB15K Pin-OutPin

Pagina 42

Alicat Scientific Innovative Flow and Pressure Solutions Alicat Digital Flow and Pressure Products: See our full product line and complete technica

Pagina 43 - Flow Vision™

Questions? Contact us at: [email protected] 520-290-6060 888-290-6060 WE GLADLY PROVIDE EXPERT TECHNICAL SUPPORT FOR YOUR APPL

Pagina 44 - Accessories

Table of Contents PageInstallation 5 Plumbing 5 Mounting 5 Application 5 Power and Signal Connections 6 RS-232 Digital O

Pagina 45 - DB15 Pin-Outs

Table of Contents Page Tare 18 Gas Select 19 Collecting Data 20 Data Format 20 Sending a Simple Script File to HyperTerminal® 21

Pagina 46

5Thank you for purchasing an M or V Series Gas Flow Meter. Please take the time to nd and read the information contained in this manual. This will he

Pagina 47 - Alicat Scientific

6Power and Signal ConnectionsPower can be supplied to your M or V Series meter through either the power jack or the 8 pin Mini-DIN connector as shown

Pagina 48 - [email protected]

7RS-232 Digital Output SignalIf you will be using the RS-232 output signal, it is necessary to connect the RS-232 Output Signal (Pin 5), the